ESP32-S2 Temperature Sensor¶
Overview¶
The ESP32-S2 has a built-in temperature sensor. The temperature sensor module contains an 8-bit Sigma-Delta ADC and a temperature offset DAC. The conversion relationship is the first columns of the table below. Among them, offset = 0 is the main measurement option, and other values are extended measurement options.
| offset | measure range(Celsius) | measure error(Celsius) |
|---|---|---|
| -2 | 50 ~ 125 | < 3 |
| -1 | 20 ~ 100 | < 2 |
| 0 | -10 ~ 80 | < 1 |
| 1 | -30 ~ 50 | < 2 |
| 2 | -40 ~ 20 | < 3 |
Application Example¶
Temperature sensor reading example: peripherals/temp_sensor.
API Reference - Normal Temp Sensor¶
Header File¶
Functions¶
-
esp_err_t
temp_sensor_set_config(temp_sensor_config_t tsens)¶ Set parameter of temperature sensor.
- Return
- ESP_OK Success
- Parameters
tsens:
-
esp_err_t
temp_sensor_get_config(temp_sensor_config_t *tsens)¶ Get parameter of temperature sensor.
- Return
- ESP_OK Success
- Parameters
tsens:
-
esp_err_t
temp_sensor_start(void)¶ Start temperature sensor measure.
- Return
- ESP_OK Success
- ESP_ERR_INVALID_ARG
-
esp_err_t
temp_sensor_read_raw(uint32_t *tsens_out)¶ Read temperature sensor raw data.
- Return
- ESP_OK Success
- ESP_ERR_INVALID_ARG
tsens_outis NULL - ESP_ERR_INVALID_STATE temperature sensor dont start
- Parameters
tsens_out: Pointer to raw data, Range: 0 ~ 255
-
esp_err_t
temp_sensor_read_celsius(float *celsius)¶ Read temperature sensor data that is converted to degrees Celsius.
- Note
- Should not be called from interrupt.
- Return
- ESP_OK Success
- ESP_ERR_INVALID_ARG ARG is NULL.
- ESP_ERR_INVALID_STATE The ambient temperature is out of range.
- Parameters
celsius: The measure output value.
Structures¶
-
struct
temp_sensor_config_t¶ Configuration for temperature sensor reading.
Public Members
-
temp_sensor_dac_offset_t
dac_offset¶ The temperature measurement range is configured with a built-in temperature offset DAC.
-
uint8_t
clk_div¶ Default: 6
-
temp_sensor_dac_offset_t
Enumerations¶
-
enum
temp_sensor_dac_offset_t¶ Values:
-
TSENS_DAC_L0= 0¶ offset = -2, measure range: 50℃ ~ 125℃, error < 3℃.
-
TSENS_DAC_L1¶ offset = -1, measure range: 20℃ ~ 100℃, error < 2℃.
-
TSENS_DAC_L2¶ offset = 0, measure range:-10℃ ~ 80℃, error < 1℃.
-
TSENS_DAC_L3¶ offset = 1, measure range:-30℃ ~ 50℃, error < 2℃.
-
TSENS_DAC_L4¶ offset = 2, measure range:-40℃ ~ 20℃, error < 3℃.
-
TSENS_DAC_MAX¶
-
TSENS_DAC_DEFAULT= TSENS_DAC_L2¶
-